How to use Avance Solo - for patients
You’ve been prescribed negative pressure wound therapy (NPWT) with Avance® Solo. On this page, you’ll find information about how to use and care for your device, as well as instructions on how to respond to different alarms.
Attention
If you experience any problems while using Avance® Solo, such as bleeding, elevated blood pressure, or issues with the device that you can’t resolve, please contact your caregiver or healthcare provider immediately.
For more detailed guidance, please see the patient user manual.What is NPWT?
NPWT is a method of treating wounds using suction. After applying the dressing, which is airtight, the pump creates a suction pressure that clears the wound of fluids. This helps to reduce swelling, protect your wound from contamination, and improve healing conditions.
How does Avance® Solo work?
Avance® Solo creates a suction pressure over your wound to help it heal. Note that the pump will make a regular pulsing sound while it is turned on. This is completely normal and tells you that it is working correctly.
When and how do I change the canister?
If the canister is full and needs to be changed, the ‘Blockage’ light will flash. The pump will also pause and emit a beeping sound. You can confirm if the canister is full by inspecting the fill level through the window on the back of the canister. If a change is required, please follow the procedure in the video below. Whenever possible, you should change the canister with your healthcare provider
May I shower while using Avance® Solo?
Yes, you are free to have light showers during treatment with Avance® Solo, but you must:
- Disconnect the pump
- Never allow jets of water to get on the dressing
Watch this video to learn what you need to do before showering to avoid damaging the device and interrupting your wound therapy.
How do I clean the device?
You may clean your Avance® Solo device by wiping it with a damp cloth or a cloth containing a small amount of non-abrasive detergent. Please do not place the pump under running water.
What should I do if the ‘Blockage’ indicator light starts to flash?
If there’s a blockage, the ‘Blockage’ indicator light will flash. The pump will also pause and emit a beeping sound. This can occur in three situations:
- There is a knot or kink in the tubing
- One of the clamps is closed
- The canister is full and needs to be changed
The following video provides brief instructions on how to fix a kink or open a clamp. For instructions on how to change the canister, please see ‘When and how do I change the canister?’
What should I do if there is a ‘Leakage’ alarm?
If there’s an air leak, the ‘Leakage’ indicator light will flash. The pump will also pause and emit a beeping sound. To learn how to deal with an air leak, please watch this video.
What type of battery does the device require?
The Avance® Solo pump requires two AA 1,5V Energizer L91 batteries. Four batteries are included with each pump.
How do I know when the batteries need to be changed?
When up to 24 hours of battery time remain, the ‘Low battery’ indication light will flash once every five seconds.
When less than 4 hours of battery time remain:
- The ‘Low battery’ indication light will stay active, flashing once every second.
- The pump will repeatedly emit an audible alarm.
How do I change the batteries?
To change the batteries, simply:
- Remove the battery compartment door on the back of the pump by sliding it upward
- Take out the two old batteries
- Insert the two new batteries, ensuring that the ‘+’ (positive) and ‘-’ (negative) symbols on each end of each battery align with the corresponding symbols in the battery compartment
- Close the battery compartment by sliding the battery compartment door downward
How do I know when therapy has ended?
When your therapy has ended, all indicator lights on the pump will flash with high intensity and the pump will emit beeping sounds in different tones. Watch this video for a demonstration.
How do I know if the device fails?
Internal failures are very rare. In the unlikely event that one should occur, all indicator lights on the device will flash with high intensity and the device will emit a beeping sound in just one tone, not in different tones, as it does when therapy has ended. If the device fails, please contact your caregiver or healthcare provider immediately.
Important points to keep in mind
-
안 함:
- 드레싱을 직접 교체
- 펌프와 캐니스터를 분리하지 않고 샤워
- 드레싱 또는 튜브를 물줄기에 노출시키거나 드레싱 또는 튜브를 물에 담그십시오.
- 분해 또는 tamp장치와 함께
- 어린이나 애완동물이 장치에 액세스할 수 있도록 합니다.
- 항공기에서 장치를 사용하십시오. 항공 여행 전에 배터리를 제거하십시오.
-
다음 사항을 기억하십시오.
- Avance® Solo를 사용하는 동안 출혈, 혈압 상승 또는 해결할 수없는 장치 문제와 같은 문제가 발생하면 즉시 간병인 또는 의료 서비스 제공자에게 문의하십시오
- 드레싱이 수축되어 만졌을 때 단단한지 확인하십시오. 그렇지 않은 경우 즉시 간병인이나 의료 제공자에게 문의하십시오
- 누액, 막힘 및 배터리 부족에 대한 시각 및 청각 경보와 이에 대응하는 방법을 숙지하십시오
-
피하다:
- 튜브가 신체 부위 주위를 꼬이도록 허용
- 튜브에 매듭과 꼬임이 생기는 것
- 벨트 클립이 허리 앞쪽에 위치하는 동안 앉아 있으면 펌프가 벨트 클립에서 밀려납니다. 앉기 전에 벨트 클립을 허리 양쪽으로 움직입니다
Related articles
-
References
- Gefen A et al. Int Wound J. 2024;21(12):e70098.